Recall Z-1378-2016 ยท LFL Instrument, Ultrasonic Surgical

Stryker Instruments Div. of Stryker Corporation

Spetzler Claw Tip, Universal - 5450-800-311 Spetzler Open Angle Micro Claw Tip, Universal - 5450-800-313 The Spetzler Claw" Tips are bone cutting tips which are a part of Strykers Neuro Spine Ear, Nose and Throat (ENT) Sonopet Portfolio. They are removable attachments intended to be used with the Sonopet Ultrasonic Handpiece.

DateApr 13, 2016 (initiated Mar 22, 2016)
StatusTerminated
ReasonThe Spetzler Claw Tip was packaged with the incorrect Tip Sleeve Assembly. The Tip Sleeve Assembly included in the packaging is the incorrect size. Risk to Health: If the Tip Sleeve Assembly does not fit the Spetzler Claw" Tip, a surgical delay may occur while an alternate Tip Sleeve Assembly is retrieved.
Root causeMixed-up of materials/components
ActionOn 3/22/2016, URGENT MEDICAL DEVICE RECALL NOTIFICATION notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions regarding this recall, please contact Stryker Instruments: Kara Spath 269-389-4518 kara.spath@stryker.com
Quantity140 units
DistributionDistributed in the states of AZ, CA, CO, CT, IL, IN, MD, MI, MN, NY, PA, SC, TN, WA, WI, and VA, and in the NETHERLANDS.