Recall Z-1278-2016 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Cios Alpha, mobile X-ray system

DateMar 31, 2016 (initiated Feb 19, 2016)
StatusTerminated
ReasonSoftware issues on Cios Alpha mobile C-Arm system
Root causeSoftware in the Use Environment
ActionSiemens sent an Customer Safety Advisory Notice on February 26, 2016, to inform them of some issues that may arise on the Cios Alpha unit. The letter identified the product the problem and the action needed to be taken by the customer. Users were instructed to change the date format to DDMMYYYY to prevent issues, not to release radiation as long as the DVD icon is blinking and to disconnect and re-connect the trolley if the system is needed immediately. In the meantime they are preparing a modification of Cios Alpha that will resolve the above-mentioned issues. As soon as the modificatio
Quantity89 systems
DistributionNationwide Distribution to the states of : AL, AR, CA, CO, CT, DE, FL, GA, IA, IL, KY, MD, MI, MN, MO, NC, NE, NJ, NV, NY, OH, OR, PA, RI, SC, TN, TX, UT, WA and WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132094CIOS ALPHA
Siemens Medi Cal Solutions, Inc., Mar 11, 2014