Recall Z-1525-2016 ยท EZL Catheter, Retention Type, Balloon

Coloplast Manufacturing US, LLC

Self-Cath Pediatric Catheter Size CH 06, Product # 5044101400, Catakig # 306. Product Usage: The Self-Cath intermittent urinary catheter discussed in this document is a pediatric size CH06 French catheter. The catheter is indicated for use by patients with chronic urine retention and patients with a post void residual volume (PVR) due to neurogenic and non-neurogenic voiding dysfunction. The cat

DateMar 22, 2016
StatusTerminated
ReasonColoplast Corp. is voluntarily recalling a single lot of Self-Cath Pediatric French Size CH 06 catheters because a small portion of the lot was identified to have missing eyelets. This recall is based upon a complaint Coloplast received regarding a catheter malfunction.
Root causeNonconforming Material/Component
ActionColoplast sent an "Urgent Medical Device Recall" letter dated March 22, 2016 to customers. The letter described the problem and the product involved in the recall. Advised consignees to immediately examine, quarantine and return the affected product. If the product has been further distributed, consignees should identify their customers and notify them of the recall. It also requested consignees to complete and return the Acknowledgement Form. For questions contact Coloplast Customer Service directly at 800-533-0464.
Quantity31,320
DistributionUS Nationwide Distribution in the states of: AL, AR, C, FL, IL KY, LA, MD, MA, MI, MN, MO, NJ,NY, NC, OK,OR, PA, TN, TX, VA and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.