Recall Z-1379-2016 ยท LWH Counter, Sponge, Surgical

Stryker Instruments Div. of Stryker Corporation

SurgiCounter scanner, part of the SurgiCount Safety-Sponge System. Model numbers 0694-097-000 and 0694-097-001. Product Usage: The SurgiCount Safety Sponge System is indicated for use in counting and recording the number of thermally labeled surgical sponges, laparotomy sponges, and towels used during surgical procedures.

DateApr 13, 2016 (initiated Mar 21, 2016)
StatusTerminated
ReasonThe affected scanner does not correctly interact with the SC360 software. When docked pre/postoperatively, the affected SurgiCounter scanner may display an error message (code=2). Also, the SurgiCounter scanner software version number is not appropriately displayed in the SC360 application during installation.
Root causeSoftware design
ActionStryker sent an URGENT MEDICAL DEVICE RECALL NOTIFICATION letters dated March 21, 2016 to affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions regarding this recall, please contact Stryker Instruments: Kelly Jo Davis 269-389-2921 kellyjo.davis@stryker.com
Quantity286
DistributionUS Nationwide Distribution in the states of AR, AZ, CA, CT, FL, GA, IN, LA, MD, MN, MO, MS, NJ, NM, NY, OH, SC, SD, TN, TX and UT.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060076SURGICOUNT SAFETY-SPONGE SYSTEM
Surgicount Medical, Mar 14, 2006