Recall Z-1390-2016 ยท GKL Counter, Cell, Automated (Particle Counter)

Becton, Dickinson and Company, BD Biosciences

BD-Retic-Count Reticulocyte Reagent System; Hematology: Used for in vitro diagnostics

DateMar 18, 2016
StatusTerminated
ReasonThe carton label for one lot of BD Retic-Count Reticulocyte Reagent System had the incorrect storage conditions. The labels on the reagent as well as the Instructions for Use are correct.
Root causeUnder Investigation by firm
ActionBD Biosciences sent an Urgent Product Recall letter on March 18, 2016, to all affected customers by certified mail. The letter identified the product the problem and the action needed to be taken by the customer. Customer Actions: 1. Examine your data files obtained using reagents from the affected lots for low performance. 2. Discard all remaining stock of BD Retic-Counter reticulocyte reagent system, Catalog Number 349204, lot number 5175559, and request replacement product. BD will replace product at no cost. 3. As a requirement of our Regulatory Tracking Process, we request that you comp
Quantity205
DistributionWorldwide Distribution - US Distribution to the states of : CT, MN, FL, NY, CA, MA, DC, UT, VA and IN., and to the countries of : Belgium, Brazil, Canada, China, India, Japan, Korea, Singapore and Taiwan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K880636RETICCOUNT SOFTWARE