ClearanceBrief
Recall Z-1310-2016 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Siemens Medical Solutions USA, Inc

AXIOM Artis (system, x-ray, angiographic) for single and biplane diagnostic imaging and interventional procedures such as cardiac and neuro angiography, general and rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.

DateFeb 19, 2016
StatusTerminated
ReasonPossible leak in the cooling system, fluids could potentially seep into the equipment cabinet. This leak can occur sporadically in affected systems.
Root causeOther
ActionSiemens mailed out an Important Customer Safety Notice to customers informing them of a corrective action that will be performed to prevent a possible leak in the detector cooling system of Axiom Artis and Artis zee systems. The notice stated the affected plastic couplings will be replaced with metal couplings to prevent fluid leaks. Siemen's service organization will contact each customer to arrange a date to perform the corrective action. Customers asked to call 1-800-888-7436 to schedule an appointment for the corrective action to be implemented.
Quantity1757 angiography systems
DistributionNationwide distribution: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and Puerto Rico
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021021MODULAR ANGIOGRAPHY SYSTEM AXIOM ARTIS
Siemens Medical Solutions USA, Inc., Jun 6, 2002