Recall Z-1596-2016 ยท LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Biomerieux Inc
VITEK 2 Gram Positive Susceptibility Test Cards (AST-P605), REF 22 325. 20 Cards/Kit. bioMerieux, Inc, Durham, North Carolina 27712. VITEK 2 is an automated system consisting of instruments, software, and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. T
| Date | May 3, 2016 (initiated Feb 24, 2016) |
| Status | Terminated |
| Reason | Product may lead to false susceptible results of Staphylococcus spp., Enterococcus spp., and S. agalactiae to Minocycline. |
| Root cause | Other |
| Action | Biomerieux, Inc. sent an Urgent Notice Product Correction on February 24, 2016, to all affected customers. The recalling firm notified their foreign subsidiary of the Field Safety Corrective Action (FSCA) by email. The local subsidiary in the country impacted has informed users impacted and provided instructions to resolve the issue. Customers were instructed to confirm the Field Safety Corrective Action letter has been distributed and reviewed by all appropriate personnel within their organization. For the referenced test kit (VITEK 2 ASTP605, Lot 485368220), refrain from reporting Minocyc |
| Quantity | 337 Kits of 20 Cards/Kit |
| Distribution | Worldwide Distribution to China Taiwan region; No US/military/government distribution. |