ClearanceBrief
Recall Z-1327-2016 ยท DZL Screw, Fixation, Intraosseous

Synthes (USA) Products LLC

Titanium Low Profile Neuro Screw, Self Drilling, 3mm, part numbers: 400.833, 400.833E Product Usage: The Low Profile Neuro Plating System is intended for use in selective trauma of the midface and craniofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.

DateApr 5, 2016 (initiated Feb 23, 2016)
StatusTerminated
ReasonOut of specification at the thread. The thread height of these screws is under-sized. The cross section of the affected area is less than that of conforming screws such that the thread height of the screws is undersized.
Root causeNonconforming Material/Component
ActionDePuy Synthes sent an Urgent Notice Medical Device Recall letter dated February 23, 2016 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were notified by mail of a potential patient impact. They were instructed to review existing inventories and call DePuy Synthes at 1800-479-6329 to obtain a Return Authorization (RA) Number and replacement order. After completing the verification section on page 4 of the letter, they were advised to return the verification section (page 3) with the product to Credit/Returns, DePuy Synthes, 1101 Sy
Quantity150 units distributed in US, 1439 units distributed OUS (foreign)
DistributionWorldwide Distribution - US Nationwide in the states of MA, VA, TN, GA, WA, IA, NJ, TN, PA, NY, CO, CT, VT, and the countries of Canada and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031807SYNTHES LOW PROFILE NEURO SYSTEM - 3 MM SCREWS
Synthes (Usa), Jul 2, 2003