Recall Z-1384-2016 ยท GXQ Dura Substitute

Synthes (USA) Products LLC

SYNTHECEL Dura Repair, 510K # K131792, K113071, Class II, Classification name: Dura Substitute Device listing number: D140596, D211012 SYNTHECEL Dura Repair is intended for use as a dura replacement for the repair of dura mater in adults.

DateDec 8, 2015
StatusTerminated
ReasonIt was reported that the SYNTHECEL Dura Repair could adhere to various anatomical structures post implantation, which may expose patients to potential risks in the event a surgeon intends to temporarily apply and subsequently remove a SYNTHECEL implant. SYNTHECEL Dura Repair labeling and promotional material are being updated to clarify that SYNTHECEL is intended to remain in place and is not intended to be removed.
Root causeLabeling False and Misleading
ActionSiemens mailed out an Urgent: Field Safety Notification letter informing customers to review the replacement package insert (GP2715-C) and discard the outdated package insert (GP2715-B) and outdated promotional materials. Customers were asked to complete the verification section (page 3 of the recall notification) and send a copy of the completed Verification Section by Fax (888)912-7351 or Scan/Email to Synthes5875@stericycle.com
Quantity200 units
DistributionNationwide: NY, PA, AL, WI, FL, MO, TN, OH, MI, MN, CO, CA, NV, AZ, LA, OR, TX, IA, IN, MS, IL, AR, WV, HI, VA, MD, WA, OK, MA, KY, NJ, CT, No foreign accounts.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113071SYNTHECEL DURA REPLACEMENT DEVICES
Synthes, Jan 9, 2012
K131792SYNTHECEL DURA REPAIR, SYNTHECEL DURA ONLAY
Synthes, Inc., Dec 16, 2013