Recall Z-1347-2016 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Medical Systems, LLC

SIGNA Pioneer MR Systems Product Usage: The SIGNA Pioneer is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or func

DateMar 1, 2016
StatusTerminated
ReasonFerrous material in the SIGNA Pioneer Table Control Box (TCB) could cause unexpected attraction to the MR system's high strength magnetic field during servicing activities. This could create a potential risk of serious injury to service personnel who attempt to inappropriately remove the TCB from the magnetic field. There have been no events reported as a result of this issue. Clinical scanni
Root causeLabeling Change Control
ActionConsignees were sent a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# 68002 dated March 1, 2016. The letter was addressed to Hospital Administrators / Risk Managers, Radiology Department Managers, and Radiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions or concerns regarding this notification,contact GE Healthcare Service at 18004371171 or your local Service Representative.
Quantity16 (2 US; 14 OUS)
DistributionWorldwide Distribution - US Nationwide in the states of SC, NC and the countries of Australia, Canada, France. Germany, Japan and Turkey.