Recall Z-1275-2016 ยท DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass

Vascular Solutions, Inc.

Guardian II Hemostasis Valve, models 8210 and 8211. The following models are only available Outside the US: FH101, FH101-T, FH101-25, and FH101-50. Product Usage: The Guardian hemostasis valves are designed to be used as conduits when interventional devices with diameters up to 8 F are inserted into the human vascular system. The devices have two seals: the low-pressure seal, and the high-

DateApr 7, 2016 (initiated Mar 4, 2016)
StatusTerminated
ReasonVascular Solutions became aware of a potential problem with the click version of the Guardian II hemostasis valves. The low pressure seal may not close properly, which may allow air to be introduced into the device and may lead to risk of an air embolism.
Root causeEmployee error
ActionConsignees were sent on 3/4/2016 via first class mail a Vascular Solutions "Urgent Medical Device Recall" letter dated March 3, 2016. The letter described the problem and the product involved in the recall. It included the "Immediate Action Required of Customer / User" which instructed the customer to identify the product, remove it from inventory, place it in a secure area, and complete and return the VSI Account Inventory Form to the Customer Service Department.
Quantity5283
DistributionNationwide Distribution in the states of AL, AR, CA, CO, DE, FL, GA, ID, IL IN, IA,KY, LA, MD, MA, MI, MO, NE,NV, NH, NY, NC, OH, PA, SD, TN, TX, UT, VA, WA, and WV. OUS: Ireland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122301GUARDIAN II HEMOSTASIS VALVE GUARDIAN II NC HEMOSTASIS VALVE
Vascular Solutions Zerusa , Ltd., Sep 7, 2012