Recall Z-2126-2016 ยท MOQ Battery, Replacement, Rechargeable

The Anspach Effort, Inc.

Battery Power Line II Battery Oscillator, DePuy Synthes Power Tools, The Anspach Effort Inc.

DateNov 25, 2015
StatusTerminated
ReasonHas the incorrect operating directions on the mode switch ring, i.e. its is labeled "FWD, LOCK, REV" instead of "ON, LOCK, ON".
Root causeLabeling design
ActionDePuy Synthes sent an Urgent Notice Recall Notification dated November 24, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. They were instructed to review their inventory and immediately remove any affected devices, complete the verification section of the letter and return it to the Anspach Effort Inc., and return any affected product to their DePuy Synthes Sales Consultant. Any questions regarding this recall / removal contact the Complain Handling Unit Manager at (561) 494-3673 or their DePuy Synthes Sales Con
Quantity20 hand pieces
DistributionUS Distribution to the state of : VA and Internationally to: Thailand and UK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.