ClearanceBrief
Recall Z-1507-2016 ยท HWC Screw, Fixation, Bone

Stryker Howmedica Osteonics Corp.

VariAx Compression Plating System

DateApr 18, 2016 (initiated Oct 10, 2013)
StatusTerminated
ReasonStryker received two reports indicating that during screw insertion under a high lateral angle a bone screw went through a oblong hole.
Root causeDevice Design
ActionStryker Orthopaedics sent notification letters/product accountability forms on October 10, 2013, via email. Customers are provided the product issue, hazards and risk mitigation along with actions needed. Customers should receive the affected products and complete the BRF and return to Stryker. Questions can be directed to Stryker. For further questions, please call (201) 831-5000.
Quantity118 units
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130009VARIAX 2 COMPRESSION PLATING SYSTEM
Stryker Trauma AG, Mar 22, 2013