Recall Z-1325-2016 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Fresenius Medical Care Renal Therapies Group, LLC

Fresenius 2008 Series Hemodialysis Delivery Systems: 2008T, 2008K2, 2008K and 2008K@home: Models: Fresenius 2008K and 2008K2 Dialysate Delivery System Fresenius 2008K@home Hemodialysis Machine with bibag System 2008T Hemodialysis Machine The Fresenius 2008 Series Machines are indicated for acute and chronic dialysis therapy

DateApr 5, 2016 (initiated Mar 10, 2016)
StatusTerminated
ReasonAdded Warning Statement: Fresenius 2008 Series Hemodialysis Machines False Blood Leak Alarm When Dialyzing Patients Treated with Hydroxocobalamin (or any form of Vitamin B-12)
Root causeDevice Design
ActionFresenius Medical Care issued an Urgent Medical Device Alert dated March 10, 2016, to all affected consignees. Consignees were instructed to place notification and Warning Addendum with their 2008 Series Hemodialysis Machine Operators Manual. A reply form was included as part of the customer communication with instructions to fill out and return via fax or scanned/emailed copy. Further information or support concerning this issue, customers were instructed to contact Fresenius Technical Services at 800-227-2572.
Quantity8384 units (8369 US) (15 OUS)
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080964FRESENIUS MODEL 2008T HEMODIALYSIS MACHINE
Fresenius Medical Care North America, Jan 14, 2009
K113427FRESENIUS 2008T HEMODIALYSIS MACHINE
Fresenius Medical Care North America, Dec 21, 2011
K120017FRESENIUS 2008 HEMODIAYSIS MACHINE WITH BIBAG SYSTEM
Fresenius Medical Care North America, Feb 2, 2012
K120505FRESENIUS 2008T HEMODIALYSIS MACHINE
Fresenius Medical Care North America, Design Cente, Mar 6, 2012
K121341FRESENIUS 2008T HEMODIALYSIS MACHINE WITH BIBAG SYSTEM
Fresenius Medical Care North America, Dec 6, 2012