Recall Z-1349-2016 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

A Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images, including digital mammography images.

DateApr 7, 2016 (initiated Jan 18, 2016)
StatusTerminated
ReasonTo inform users about the possible incorrect values for Distance Measurements when using certain modalities in combination with syngo Imaging.
Root causeSoftware design
ActionSiemens mailed a Customer Advisory Notice to their customers instructing them to check their modalities for the usage of the mentioned DICOM attributes with different values. They recommended upgrading the workplace(s) to syngo Imaging Studio version VB36D if they have such modalities and it cannot be ensured by the modality provider that only one attribute is filled or both attributes are filled with the same value. Customers were asked to contact their local service organization to clarify the compatibility of their syngo Imaging server version with workplace version VB36D.
Quantity49 imaging systems
DistributionUS (nationwide) including the states of AL, CA, CO, CT, FL, IA, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OK, OR, PA, TN, TX, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071114SYNGO IMAGING (VERSION V30)
Siemens AG Medical Solutions, May 22, 2007
K081734SYNGO IMAGING (VERSION V31)
Siemens AG Medical Solutions, Jul 3, 2008