Recall Z-1270-2016 ยท HQF Laser, Ophthalmic

Laserex Systems Inc.

Ellex Medical Fixed eye Safety filters Model Integre S LP5532. The filter is located in the binocular mount of the laser delivery head. The filter prevents transmission of treatment laser light to physician eyes while the aiming beam to remain visible. .

DateFeb 1, 2016
StatusTerminated
ReasonThe Firm discovered that there is a design error which is traceable back to 2012 when a design change on the safety filter optic adaptor was implemented.
Root causeDevice Design
ActionEllex Medical Pty Ltd. Planned Action to bring into Compliance: 1.Field Corrective Action replacing all affected safety filters of all units manufactured between July 2012 and February 2016, 2.All repairs will be made at no cost to the purchaser; and 3.The corrections will be completed by March 30, 2017. CDRH approves the CAP subject to the following conditions: Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. For further questions please call (800) 824-7444.
Quantity72 units plus 2 spare units
DistributionWorldwide Distribution - US Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052777LASEREX DUO FAMILY OF DUAL WAVELENGTH OPHTHALMIC LASERS INCLUDING MODELS INTEGRE DUO LP1RG-S, LP1RG-D
Ellex Medical Pty, Ltd., Mar 16, 2006