ClearanceBrief
Recall Z-1064-2017 ยท KXJ Table, Radiologic

Villa Radiology Systems LLC

Apollo: Models: 9784000131, 9784000231, 9784020131, 9784020231, 9784100231, 9784120131, 9784120231, 9784130131 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.

DateFeb 19, 2016
StatusOpen, Classified
ReasonThe analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.
Root causeDevice Design
ActionConsignees were notified of the recall via letter starting on 2/19/16. The letter explains the issue and actions to be taken by the consignee and manufacturer.
Quantity66 distributed in US
DistributionAL AZ CA CO CT DC FL IA IL IN LA MA MD MI MN MO MS NJ NY OH PA Puerto Rico TN TX UT WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050190APOLLO
Villa Sistemi Medicali S.P.A., Mar 25, 2005