Recall Z-1268-2016 ยท EAZ Light, Operating, Dental

Pelton & Crane / Marus / DCI Equipment / KaVo

Helios HL3T and HLT Track Mounted Dental Lights Product Usage: The product is intended for use by dental professionals to illuminate the oral cavity while performing examinations and dental procedures.

DateMar 8, 2016
StatusTerminated
ReasonThe firm received two customer complaints local Pelton & Crane distributors that the Helios 3000 track mounted dental lights were missing the two 8/32x3/16 set screws during the initial installation of the dental light.
Root causeNonconforming Material/Component
ActionPelton & Crane sent an Urgent Medical Device Recall letters to affected customers on March 8, 2016 and will be completed by 3/11/2016. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete and return the Field Correction Acknowledgement Form in the enclosed self-address stamped envelope to Pelton & Crane. For questions or concerns regarding this recall please call 1-800-659-6560, ext. 7227.
Quantity156 units
DistributionUS Nationwide Distribution in the states of AZ, AR, CA, CO, CT, FL, IA, ID, IL, IN, MA, MI, MN, NJ, NY, NC, OH, OR, PA, TN, TX, UT, VA, WI and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.