Recall Z-1633-2016 ยท NRP Sirolimus Test System

Siemens Healthcare Diagnostics, Inc.

Dimension clinical chemistry system, Sirolimus (SIRO) Flex reagent cartridge (DF306 SMN 10464331) Lot EB6064, Open Well Instability The SIRO method is an in vitro diagnostic test for the quantitative measurement of sirolimus in whole blood on the Dimension clinical chemistry system.

DateFeb 9, 2016
StatusTerminated
ReasonSiemens Healthcare Diagnostics has received customer complaints regarding QC shifts and imprecision after one day when using Dimension SIRO lot EB6064. They have confirmed SIRO lot # EB6064 does not meet the open well stability claim of 2 days.
Root causeUnder Investigation by firm
ActionSiemens sent an "URGENT MEDICAL DEVICE RECALL" letter dated February 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Actions to be Taken by the Customer: Discontinue use of and discard the kit lot listed in Table 1. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the product listed in Table 1, immediately contact your
Quantity394
DistributionWorldwide distribution and US including Puerto Rico and to the states of : TX, FL, CA, LA, DC, GA, IL, AR, WI, MI and MD.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081857DIMENSION SIRO FLEX REAGENT CARTRIDGE AND DIMENSION SIROLIMUS CALBIRATOR, MODELS DR 306, DC 306
Siemens Healthcare Diagnostics, Inc., Oct 30, 2008