Recall Z-1565-2016 ยท FKX System, Peritoneal, Automatic Delivery

Baxter Healthcare Corp.

Automated peritoneal dialysis (APD) cycler

DateFeb 16, 2016
StatusTerminated
ReasonBaxter Healthcare Corporation is sending this communication to inform you of incomplete instructions in the AMIA Automated Peritoneal Dialysis (PO) System Clinician Guide for calculating the total recommended solution therapy volume. Specifically, the instructions do not specify the need for an extra 200 mL of PD solution in order to prime the patient line and for air purge operations. The tot
Root causeUnder Investigation by firm
ActionAn Urgent Device Correction communication was sent to the one affected customer via U.S.P.S., first class mail on 02/11/2016. A Clinician Guide Addendum was created to instruct clinicians of the extra 200 mL of PD solution required to prime the patient line and for air purge operations. This Addendum was provided as an attachment to the Urgent Device Correction communication letter.
Quantity10 units
DistributionUS Consignee: NY ***Foreign Consignee: Ecuador
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151525Amia Automated PD System with Sharesource
Deka Research and Development, Oct 9, 2015