Recall Z-1098-2016 ยท LNH System, Nuclear Magnetic Resonance Imaging

Deerfield Imaging

Intra operative MRI Systems: Neuro II-SE , Neuro III-SV, iMRI 1.5T A, and iMRI 3T S. Indicated for use as a magnetic resonance diagnostic device (MRDD).

DateMar 11, 2016 (initiated Feb 10, 2016)
StatusTerminated
ReasonEmergency helium venting lines (i.e. quench lines) may not have been properly installed.
Root causeDevice Design
ActionConsignees were sent a IMRIS "Urgent Field Safety Notice" letter dated February 10, 2016. The letter described the problem and the product involved in the recall. The firm also stated that it is important that customers review the portion of the quench line vent described and ensure that it is installed to the specifications supplied as part of the Planning Guide. Customers were reminded that it is important to have an emergency plan in place in the event Helium gas escapes into the magnet room or any other area. For questions customers can contact their local customer support hotline or
Quantity35 USA, 14 OUS
DistributionWorldwide Distribution -- US, including the states of CA, CT, FL, GA, IL, IN, MD, MA, MN, MO, NH, NC, OH, OR, PA, TX, UT, VA, and WI; and ,countries of CHINA, CANADA, AUSTRALIA, FRANCE, JAPAN, and GERMANY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061916NEURO II-SE, MODELS OR-MB-DR AND OR-DR
Imris, Inc., Aug 11, 2006
K071099NEURO II-SE, MODELS: OR-DR-OR, OR-MB
Imris, Inc., May 22, 2007
K083137NEURO III-SV
Imris, Inc., Dec 16, 2008
K133692IMRI 1.5T AND IMRI 3T S
Imris, Inc., Feb 10, 2014