Recall Z-1639-2016 ยท LTW Susceptibility Test Cards, Antimicrobial

Biomerieux Inc

Vitek 2 Gram Positive Susceptibility Test Cards (AST-P636 REF 417951), 20 cards/kit. Product Usage: The VITEK 2 Gram Positive Susceptibility Card is intended for use with VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed in the Product Informatio

DateMay 11, 2016 (initiated Feb 24, 2016)
StatusTerminated
ReasonProduct insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.
Root causeLabeling Change Control
ActionBiomerieux sent an Urgent Product Correction Notice letter via email on February 24, 2016 to affected customers . The letter identified the affected product, problem and actions to be taken. Customers were asked to complete the attached Acknowledgement Form (Attachment A) and return it to Biomerieux, Inc. For questions contact your local bioMrieux representative.
Quantity495 kits
DistributionForeign distribution in the countries of Czech Republic, Hong Kong (China), Hungary, Japan, Poland, South Africa, Switzerland, and Turkey.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.