ClearanceBrief
Recall Z-1338-2016 ยท MDS Sensor, Glucose, Invasive

Dexcom Inc

Dexcom G4 PLATINUM (Pediatric) Receiver with Share Receiver Part Number: MT22608 Receiver (mg/dL), Black (pediatrics); MT22608-PNK, Receiver (mg/dL), Pink (pediatrics); MT22608-BLU, Receiver (mg/dL), Blue (pediatrics). Intended to detect trends and track glucose patterns in persons with diabetes.

DateApr 11, 2016 (initiated Feb 23, 2016)
StatusTerminated
ReasonCustomers may not receive an intended audible alarm or alert if relying on hearing the alarm or alert.
Root causeDevice Design
ActionDexcom sent out a customer notification letter by certified mail to all direct customers on 02/23/16 notifying customers that they may not receive an intended audible alarm or alert if relying on hearing the alarm or alert. As a result customers may not detect a severe hypoglycemic (low glucose) or hyperglycemic (high glucose) event. The letter instructs the customer to peridically test the alarms and alerts on their receiver to make sure that the alarms and alerts are functioning properly (especially if the receiver gets wet of is dropped). Receiver part numbers and instructions for testin
Quantity19,586
DistributionWorldwide Distribution-US (nationwide), Canada, Italy, Netherlands, Belgium, Norway, New Zealand, Poland, Sweden, Denmark, Slovenia, Turkey, South Africa, India, Israel, Hungary, Hong Kong, UK, Ireland, France, Finland, Spain, Portugal, Colombia, Chile, Germany, Austria, Switzerland, Kuwait, United
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.