Recall Z-1184-2016 ยท INK Stretcher, Wheeled, Powered
Stryker Medical Division of Stryker Corporation
Prime Series Stretchers (Fifth Wheel, Big Wheel and Zoom) Powered stretcher
| Date | Feb 5, 2016 |
| Status | Terminated |
| Reason | The hydraulic jack assemblies (PN: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) were manufactured using the incorrect oil. The oil used does not contain the same wear resistance additives as the specified oils and therefore its use increases the potential for wear which may cause a leak event or jack drift. Testing was done to evaluate the jacks with incorrect oil and leak events we |
| Root cause | Mixed-up of materials/components |
| Action | On 2/5/2016, URGENT MEDICAL DEVICE NOTIFICATION notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If you have any urgent questions or concerns, please, contact Stryker Customer Service at (1-800-327-0770). Our normal business hours are Monday-Friday 8 a.m.-5 p.m. (ET). |
| Quantity | 36 |
| Distribution | Distributed in DC and the states of AL, CA, FL, MN, MT, NC, NE, NY, PA, RI, TN, TX, UT, and WV. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K140095 | PRIME ZOOM STRETCHER Stryker Corporation, May 20, 2014 |