Recall Z-1188-2016 ยท FWE Camera, Television, Microsurgical, With Audio

Richard Wolf Medical Instruments Corp.

The Vesa 75/100 adaptation with rotation is an accessory to a surgical camera intended to be used to record operative procedures.

DateFeb 29, 2016
StatusTerminated
ReasonSupplier of the "VESA 75 / 100 adaptation with rotation", which is part of their monitor mounts and support arms may be missing riveting.
Root causeProcess control
ActionRICHARD WOLF sent an Urgent Safety Information letter dated March 2, 2016, to all affected consignees. The letter described the nature of the recall regarding the monitor holding plate (VESA 75/100 mount) in our Flexion-Port 300, 350 and 400 swivel arm product series. According to the firm insufficient riveting of the rotation elements the rivets may loosen during continuous use (e.g. when adjusting the tilt angle) and in an extremely unlikely event the monitor holding plate (VESA 75/100 mount) may loosen. According to Richard Wolf this can result in a potential danger for the patient, user a
Quantity108 units
DistributionNationwide Distribution including Virginia, Pennsylvania, Florida, California, Texas, North Carolina and Illinois.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.