Recall Z-1324-2016 ยท FJI Dialyzer, Capillary, Hollow Fiber

Fresenius Medical Care Renal Therapies Group, LLC

Fresenius Hemoflow F3 Dialyzer Part Number: 0520165A Indicated for Acute and Chronic Hemodialysis

DateApr 5, 2016 (initiated Feb 26, 2016)
StatusTerminated
ReasonDialyzer header leak due to possible improper torque
Root causeProcess control
ActionThe firm, Fresenius Medical Care Renal Therapies Group, LLC issued an "URGENT MEDICAL DEVICE RECALL" letter dated 2/24/16 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use of your Hemoflow F3 Dialyzer; place all affected units in a secure, segregated area; contact your FMCRTG, LLC Customer Service Team at 1-800-323-5188 for instructions on how to return the recalled product, and fill out and return the attached Reply Form via fax to: 781-699-9796 ATTN: Quality Department or scan and email to: NOTIF
Quantity1,650 units
DistributionUS Nationwide Distribution including states of: AZ, CA, CT, DC, GA, IL, IN, MI, MN, MO, MS, NY, OH, PA, SC, SD, TX, and WA.