Recall Z-1236-2016 ยท BTO Tube, Tracheostomy (W/Wo Connector)

Ambu Inc.

Ambu, Aura Gain - Single Use Laryngeal Mask- Sterile for use as an alternative to a face mask for achieving and maintaining control of the airway during routine emergency anesthetic procedures.

DateMar 24, 2016 (initiated Feb 10, 2016)
StatusTerminated
ReasonAmbu, Aura Gain laryngeal Mask, was found to have lack of Sterility Assurance. The firm discovered the problem during final quality control at manufacturing site. Damage to the sterile barrier (the pouch) may compromise a medical device's sterility,leading to contamination of the product and,ultimately,patient complications.
Root causeProcess control
ActionFiled Safety Notice was sent via USPS Certified Mail to all customers on Dec 10,2015.
QuantityTotal Numbers of unit distributed: 45990 pes (USA)
DistributionPlease see attached consignee list
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.