Recall Z-1024-2016 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips Electronics North America Corporation

SENSE Body Coil 1.5T with identification; used in the ACS NT, Intera 1.5T and Achieva 1.5T MR system

DateFeb 26, 2016 (initiated Jul 6, 2015)
StatusTerminated
ReasonWrong positioning of the coil cables can create unintended resonances via the patient or elevated coil cable temperature due to coupling to the QBC. The combined use of the Sense coil, high SAR level scanning and placement of the cables contrary to instructions for use can lead to patient burns.
Root causeDevice Design
ActionOn 6/18/2015, the Field Safety Notice / Medical Device Correction Notice was sent to customers via certified mailing. The letter explains the issue and provides user instructions. A sleeve is being developed that will be applied to the cable on the Sense Body Coil 1.5T and provide consistent spacing between the cable and the patient, thereby reducing the chance of RF interaction and heating. This sleeve will be distributed as part of a Field Change Order with reference FCO781 00437 for Sense Body Coils in the installed base and will be installed by the Philips Field Service Engineers.
Quantity5672
DistributionUS: Nationwide OUS: Canada, Afghanistan Albania Algeria Andorra Argentina Australia Austria Azerbaijan Bahamas Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Botswana Brazil Burkina Faso Chile China Colombia Columbia Congo, Democratic Costa Rica Croatia Cyprus Czech Republic Denmark Domin
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001796GYROSCAN INTERA (R7.5)
Philips Medical Systems North America, Inc., Sep 6, 2000
K013344INTERA I/T
Philips Medical Systems North America Co., Dec 11, 2001
K030520INTERA FAMILY
Philips Medical Systems North America Co., Mar 6, 2003
K945945GYROSCAN ACS-NT, GYROSCAN T5-NT AND GYROSCAN T10-NT
Philips Medical Systems North America, Inc., Dec 22, 1995