Recall Z-1163-2016 ยท KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
Zimmer Manufacturing B.V.
Product 65 consists of all product under product code: KTT and same usage: Item no: 119413594 VERSA-FX II KEYLESS SHORT 119413595 VERSA-FX II KEYLESS SHORT 119413596 VERSA-FX II KEYLESS SHORT 119413004 VERSA-FX II KEYLESS TUBE 119413005 VERSA-FX II KEYLESS TUBE 119413006 VERSA-FX II KEYLESS TUBE 119413008 VERSA-FX II KEYLESS TUBE 119413504 VERSA-FX II KEYLESS TUBE 119413505 VERSA-FX II KEYLESS TU
| Date | Mar 12, 2016 (initiated Jan 11, 2016) |
| Status | Terminated |
| Reason | As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile. |
| Root cause | Packaging |
| Action | Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached |
| Quantity | 5538 |
| Distribution | US Nationwide Distribution |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K150818 | Versa-Fx Femoral Fixation System, Versa-Fx II Femoral Fixation System, Free-Lock Femoral Fixation System Zimmer, Inc., May 1, 2015 |