ClearanceBrief
Recall Z-1161-2016 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Zimmer Manufacturing B.V.

Product 63 consists of all product under product code: LPH and same usage: Item no: 620004020 TRILOGY ACET SHELL 40MM O 620004220 TRILOGY ACET SHELL 42MM O 620004420 TRILOGY ACET SHELL 44MM O 620004620 TRILOGY ACET SHELL 46MM O 620004621 TRILOGY ACET SHELL 46MM O 620004820 TRILOGY ACET SHELL 48MM O 620004822 TRILOGY ACET SHELL 48MM O 620004823 TRILOGY ACET SHELL 48MM O 620005020 TRILOGY ACET SHEL

DateMar 12, 2016 (initiated Jan 11, 2016)
StatusTerminated
ReasonAs a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Root causePackaging
ActionZimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached
Quantity2063
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K934765TRILOGY ACETABULAR SYSTEM
Zimmer, Inc., Apr 29, 1994