Recall Z-1156-2016 ยท HRS Plate, Fixation, Bone

Zimmer Manufacturing B.V.

Product 58 consists of all product under product code: HRS and same usage: Item no: 234800904 PROXIMAL DORSAL ULNA PLT 234800906 PROXIMAL DORSAL ULNA PLT 234800908 PROXIMAL DORSAL ULNA PLT 234800910 PROXIMAL DORSAL ULNA PLT 234801004 PROXIMAL DORSAL ULNA PLT 234801006 PROXIMAL DORSAL ULNA PLT 234801008 PROXIMAL DORSAL ULNA PLT 234801010 PROXIMAL DORSAL ULNA PLT 10H LT 234701304 PROXIMAL MEDIAL TI

DateMar 12, 2016 (initiated Jan 11, 2016)
StatusTerminated
ReasonAs a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Root causePackaging
ActionZimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached
Quantity119
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150121Zimmer Periarticular Plating System
Zimmer, Inc., Feb 23, 2015