ClearanceBrief
Recall Z-1155-2016 ยท JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented

Zimmer Manufacturing B.V.

Product 57 consists of all product code: JDI and same usage: Item no: 763401000 PRESSFIT FEM ST FX 10 X 115 763401100 PRESSFIT FEM ST FX 11 X 1 763401200 PRESSFIT FEM ST FX 12 X 1 763401300 PRESSFIT FEM ST FX 13 X 1 763401400 PRESSFIT FEM ST FX 14 X 1 763401500 PRESSFIT FEM ST FX 15 X 1 For use in total or hemi hip arthroplasty

DateMar 12, 2016 (initiated Jan 11, 2016)
StatusTerminated
ReasonAs a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Root causePackaging
ActionZimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached
Quantity0
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.