ClearanceBrief
Recall Z-1152-2016 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Zimmer Manufacturing B.V.

Product 54 consists of all product under product code: JWH and same usage: Item no: 596009900 NEXGEN TAPER PLUG for use in total knee arthroplasty

DateMar 12, 2016 (initiated Jan 11, 2016)
StatusTerminated
ReasonAs a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Root causePackaging
ActionZimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached
Quantity1767
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K991581NEXGEN COMPLETE KNEE SOLUTION LEGACY POSTERIOR STABILIZED (LPS); LPS-FLEX FIXED BEARING FEMORAL AND ARTICULAR SURFACE CO
Zimmer, Inc., Jul 30, 1999