Recall Z-1245-2016 ยท HKN Refractor, Manual, Non-Powered, Including Phoropter

Nidek Inc

RT-5100 (Epic 5100) Refractor and RT 3100 Refractor: Product Usage: Instrument providing means of positioning spherical and cylindrical lenses, prisms and other optical devices in front of a subjects eye for purpose of determining refractive error and binocular functions.

DateMar 24, 2016 (initiated Feb 29, 2016)
StatusTerminated
ReasonThere has been a reoccurrence of an issue on the RT 5100 and RT 3100 Refractor of the Near Point Chart Arm lowering spontaneously with potential for slight injury to face or nose of patient.
Root causeComponent design/selection
ActionNidek sent a Recall letters dated February 29, 2016 to all affected site. The letter identified the affected product, problem and actions to be taken. Nidek Customer Service will be replacing the affected product with new parts to address this recall. For questions call Customer Service at 1-800-722-0219.
Quantity2703 separate serial numbers
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.