Recall Z-1088-2016 ยท OWB Interventional Fluoroscopic X-Ray System
Medtronic Navigation, Inc.
Medtronic 0-arm 02 Imaging System Catalog Number: Bl-700-02000 The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm O2 Imaging System is compatible with certa
| Date | Mar 8, 2016 (initiated Jan 29, 2016) |
| Status | Terminated |
| Reason | Screws holding the detector panel in as a result the detector panel could make contact with other parts with in the 0-arm 02 Imaging System, causing noise and vibration when being positioned or during 3D Imaging could lead to poor image quality |
| Root cause | Process control |
| Action | Medtronic Navigation issued a Field Safety Notice dated January 29, 2016, to notify them of a hardware correction in the unit. A Medtronic service representative will service the system and arrange for an inspection of the system. This representative will properly torque the eight (8) screws if they find them to be loose. In the meantime, if noise or vibration are observed during use of the system customers should contact their Medtronic representative to report the issue. Customers with questions were instructed to call 720-890-3160. For questions regarding this recall call 978-698-6000. |
| Quantity | 1 unit |
| Distribution | Worldwide Distribution to Austria only. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K151000 | O-Arm O2 Imaging System Medtronic, Inc., Aug 6, 2015 |