ClearanceBrief
Recall Z-1142-2016 ยท HWC Screw, Fixation, Bone

Zimmer Manufacturing B.V.

Product 44 consists of all product under product code: HWC and same usage: Item no: 114607099 MAGMA-FX CANN SCREW 7.0MM 114604099 MAGNA FX CANN SCREW 7.0MM 114603000 MAGNA-FX CANN SCREW 7.0MM 114603500 MAGNA-FX CANN SCREW 7.0MM 114604000 MAGNA-FX CANN SCREW 7.0MM 114604032 MAGNA-FX CANN SCREW 7.0MM 114604500 MAGNA-FX CANN SCREW 7.0MM 114604532 MAGNA-FX CANN SCREW 7.0MM 114604599 MAGNA-FX CANN SC

DateMar 12, 2016 (initiated Jan 11, 2016)
StatusTerminated
ReasonAs a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Root causePackaging
ActionZimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached
Quantity4335
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142442MAGNA-FX AND MINI MAGNA-FX CANNULATED SCREW FIXATION SYSTEM
Zimmer, Inc., Oct 3, 2014