ClearanceBrief
Recall Z-1130-2016 ยท HWC Screw, Fixation, Bone

Zimmer Manufacturing B.V.

Product 32 consists of all product under product code: HWC and same usage: Item no: 47115401200 HERBERT BONE SCREW, 3.0MM 47115401400 HERBERT BONE SCREW, 3.0MM 47115401600 HERBERT BONE SCREW, 3.0MM 47115401800 HERBERT BONE SCREW, 3.0MM 47115402000 HERBERT BONE SCREW, 3.0MM 47115402200 HERBERT BONE SCREW, 3.0MM 47115402400 HERBERT BONE SCREW, 3.0MM 47115402600 HERBERT BONE SCREW, 3.0MM 4711540280

DateMar 12, 2016 (initiated Jan 11, 2016)
StatusTerminated
ReasonAs a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Root causePackaging
ActionZimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached
Quantity4018
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143165Herbert/Whipple Bone Screw System, Herbert/Whipple Bone Screw System, Herbert Cannulated Bone Screw System, Herbert Mini
Zimmer, Inc., Mar 31, 2015