Recall Z-0826-2016 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

software for Syngo Dynamics a Picture Archiving and Communication System (PACS) intended for acceptance, transfer, display, storage, archive and manipulation of digital medical images, including quantification and report generation. Syngo Dynamics is not intended to be used for reading of mammography images.

DateFeb 19, 2016 (initiated Jan 29, 2016)
StatusTerminated
ReasonSiemens is releasing a software update that addresses an issue of mixing data from multiple patients. In rare situations, echo trend graphs may mix data from multiple patients.
Root causeSoftware design
ActionThe firm, Siemens Healthcare, sent a "Customer Advisory Notice" dated 1/19/2019 to its customers.The notice described the product, problem and actions to be taken. The customers were instructed to create a new department if immediate resolution is necessary, use workplace for reporting and hand over the customer letter before or during installation of field modification(PSL) distributed with Update Instruction SY013/16P. A software update to resolve reported issues is being released via an Update Instruction SY013/16/P. If you have any questions, contact Regulatory Technical Specialist at 61
Quantity151
DistributionUS Nationwide Distribution including states of: AL AR AZ CA CO CT DE FL GA IL IN KS LA MA MD MI MN MO NC ND NE NJ NM NY OH OR PA SC TN TX UT VA WA and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023772KINETDX PICTURE ARCHIVING AND COMMUNICATIONS SYSTEM
Acuson Corp., Nov 22, 2002
K102150SYNGO DYNAMICS, VERSION 9.0
Siemens Medical Solutions USA, Inc., Oct 5, 2010