Product 23 consists of all product code: JDI and same usage: Item no: 801100220 CPT 2 HIP STEM, X-OFFSET 801100320 CPT 3 HIP STEM, X-OFFSET 801100420 CPT 4 HIP STEM, X-OFFSET 801100005 CPT HIP STEM PETITE 801100700 CPT HIP STEM SZ 2 LONG 801100800 CPT HIP STEM SZ 3 LONG 801100222 CPT MOD HD ST/ST 22MED SH 801100322 CPT MOD HD ST/ST 22MMD LN 801100122 CPT MOD HD ST/ST 22MMD SH 801100126 CPT MOD HD
| Date | Mar 12, 2016 (initiated Jan 11, 2016) |
| Status | Terminated |
| Reason | As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile. |
| Root cause | Packaging |
| Action | Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached |
| Quantity | 10 |
| Distribution | US Nationwide Distribution |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.