ClearanceBrief
Recall Z-1120-2016 ยท JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented

Zimmer Manufacturing B.V.

Product 22 consists of all product under product code: JDI and same usage: Item no: 811400218 CPT 12/14 COCR REVISION S 811400318 CPT 12/14 COCR REVISION S 811400420 CPT 12/14 COCR REVISION S 811400423 CPT 12/14 COCR REVISION S 811400426 CPT 12/14 COCR REVISION S 811400010 CPT 12/14 COCR SIZE 0 EXT 811400000 CPT 12/14 COCR SIZE 0 STD 811400110 CPT 12/14 COCR SIZE 1 EXT 811400100 CPT 12/14 COCR

DateMar 12, 2016 (initiated Jan 11, 2016)
StatusTerminated
ReasonAs a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Root causePackaging
ActionZimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached
Quantity16
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030265CPT 12/14 HIP PROSTHESES, MODEL 00-8114-040/050-00
Zimmer, Inc., Mar 4, 2003