Recall Z-1118-2016 ยท KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component

Zimmer Manufacturing B.V.

Product 20 consists of all prod oct under poduct code: KTT and same usage: Item no: 47116201900 COMPRESSION SCREW 1/2IN 47116202300 COMPRESSION SCREW 1-1/2IN 47116202200 COMPRESSION SCREW 1-1/4IN 47116202400 COMPRESSION SCREW 1-3/4IN 47116202100 COMPRESSION SCREW 1IN 47116202000 COMPRESSION SCREW 3/4IN Product Usage: Single/multiple component metallic bone fixation appliances and accessories ar

DateMar 12, 2016 (initiated Jan 11, 2016)
StatusTerminated
ReasonAs a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Root causePackaging
ActionZimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached
Quantity3415
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150818Versa-Fx Femoral Fixation System, Versa-Fx II Femoral Fixation System, Free-Lock Femoral Fixation System
Zimmer, Inc., May 1, 2015