Recall Z-1115-2016 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Zimmer Manufacturing B.V.

Product 17 consists of all product under product code: LPH and same usage: Item no: 625004515 BONE SCREW 4.5X15 SELF-TA 625004520 BONE SCREW 4.5X20 SELF-TA 625004525 BONE SCREW 4.5X25 SELF-TA 625004530 BONE SCREW 4.5X30 SELF-TA 625004535 BONE SCREW 4.5X35 SELF-TA 625004540 BONE SCREW 4.5X40 SELF-TA 625004550 BONE SCREW 4.5X50 SELF-TA 625006515 BONE SCREW 6.5X15 SELF-TA 625006520 BONE SCREW 6.5X

DateMar 12, 2016 (initiated Jan 11, 2016)
StatusTerminated
ReasonAs a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Root causePackaging
ActionZimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached
Quantity87341
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K934765TRILOGY ACETABULAR SYSTEM
Zimmer, Inc., Apr 29, 1994