Recall Z-1109-2016 ยท JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
Zimmer Manufacturing B.V.
Product 11 consists of all product under product code JDI , and same usage: Item no: 902602500 6 DEGREE COCR FEM HEAD 26 902602600 6 DEGREE COCR FEM HEAD 26 902602700 6 DEGREE COCR FEM HEAD 26 902602800 6 DEGREE COCR FEM HEAD 28 902602900 6 DEGREE COCR FEM HEAD 28 902603000 6 DEGREE COCR FEM HEAD 28 902603300 6 DEGREE COCR FEM HEAD 32 902603200 6 DEGREE COCR FEM HEAD 32 Product Usage: For use i
| Date | Mar 12, 2016 (initiated Jan 11, 2016) |
| Status | Terminated |
| Reason | As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile. |
| Root cause | Packaging |
| Action | Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached |
| Quantity | 366 |
| Distribution | US Nationwide Distribution |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K960278 | COBALT-CHROMIUM-MOLYBDENUM (CO-CR-MO) ALLOY FEMORAL HEADS W/CO-NIDIUM SURFACE TREATMENT Zimmer, Inc., Jul 25, 1996 |