Recall Z-2353-2016 ยท FNL Bed, Ac-Powered Adjustable Hospital

Stryker Medical Division of Stryker Corporation

S3 MedSurg Bed, Model 3005. Stryker Medical

DateAug 8, 2016 (initiated Feb 2, 2016)
StatusTerminated
ReasonCPU board failures cause fowler (backrest) electronic controls to stop working.
Root causeComponent change control
ActionStryker Medical initiated a voluntary recall of the S3 MedSurg Bed, Model 3005, with notices sent on February 2, 2016 via certified mail. Stryker medical identified that Fowler does not raise up to desired angle; and the Fowler does not raise electronically, but manual function remains operational through the manual CPR release. Actions Needed for Customers: 1. Locate the bed(s) listed by serial number on the attached business reply form. 2. A Stryker ProCare Representative will contact your facility to schedule a time for repair. 3. Fax (269)488-8691 or email productfieldaction@stryker.com t
Quantity8490
DistributionDistribution US nationwide, Canada, and Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.