Recall Z-1032-2016 ยท GCJ Laparoscope, General & Plastic Surgery

Novadaq Technologies Inc.

PINPOINT Endoscopic Fluorescence Imaging System

DateJan 29, 2016
StatusTerminated
ReasonIt has been determined that the PINPOINT Operator's Manual does not contain validated parameters for gravity-cycle autoclave sterilization of the PINPOINT laparoscopes and light guide cable in the reprocessing tray.
Root causeProcess design
ActionNovadaq sent an Urgent Device Correction letter dated February 2, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Recipients of this notice have received shipments that include the inaccurate Operator's Manual. Please follow these instructions: 1. Discard any old PINPOINT Operator's Manuals Rev J through M. 2. Accept and review the new PINPOINT Operator's Manual (Rev N). 3. Complete and sign the enclosed "Acknowledgement of Receipt" form, scan it and email to PI NPOINT@novadaq.com. Any questions or concer
Quantity110 units
DistributionWorldwide Distribution - Nationwide Distribution including Puerto Rico and to the countries of : Bangladesh, Canada, Hong Kong, India, and Israel.