ClearanceBrief
Recall Z-1039-2016 ยท GDF Guide, Needle, Surgical

Stryker Instruments Div. of Stryker Corporation

50000 (Inrad) AccuPlace Drace Stereotaxic Needle Guide; Assists the user with inserting a needle at a predetermined angle for computer tomography.

DateMar 3, 2016 (initiated Feb 4, 2016)
StatusTerminated
ReasonDuring an evaluation of the product packaging, it was determined that there is a potential for the product to cause damage to the packaging, creating a possible breach in sterility.
Root causePackage design/selection
ActionStryker sent URGENT MEDICAL DEVICE RECALL NOTIFICATIONS to Customers via Overnight service. Stryker Sales Representatives were notified via email. 1. Immediately review this Recall Notification. 2. Check all stock areas and/or operating room storage to determine how many AccuPlace" Needle Guides are at your facility. Discontinue use of the AccuPlace" Needle Guides. 3. Complete the enclosed Business Reply Form (BRF) to confirm receipt of this notification and identify how many, if any, affected items are currently in your inventory. Please complete and return the BRF even if you dont have any
Quantity5,800 units total
DistributionNationwide Distribution-including the states of AZ, CA, CO, CT, FL, GA, IA, KS, MA, MI, MN, MO, NY, OH, OK, PA, TX, VA, WA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K904172ACCUPLACE (TM)
Dlp, Inc., Nov 21, 1990