Recall Z-0854-2016 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Philips Electronics North America Corporation
Philips IntelliVue Module Measurement X2 Model: M3002A
| Date | Jan 4, 2016 |
| Status | Terminated |
| Reason | The ST elevation alarm on the Patient Monitor or standalone X2 Measurement Module will not sound when indicated for all chest leads derived using Hexad 12-Lead ECG Monitoring in the Host Monitor |
| Root cause | Software Design Change |
| Action | Philips Healthcare issued the Field Safety Notice on January 4, 2016, and informed customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Field Safety Notice: Until your software is upgraded, please make sure that the ST Analysis is switched ON when using STE measurement in the Hexad 12-lead ECG Monitoring. This can be done by entering the ST Analysis Menu and se |
| Quantity | 13,300 |
| Distribution | Worldwide Distribution - Nationwide Distribution and to the countries of : Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Dominican Republic, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K150310 | IntelliVue Patient Monitor Philips Medizin Systeme Boeblingen GmbH, May 13, 2015 |