ClearanceBrief
Recall Z-1437-2016 ยท KKX Drape, Surgical, Antimicrobial

Customed, Inc

Catalog Number: 9001771 MAYO STAND COVER STERILE Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc.

DateApr 14, 2016 (initiated Jan 7, 2016)
StatusTerminated
ReasonRecall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
Root causePackaging
ActionRecall Notification Letter sent to direct accounts via certified mail with return receipt on January 7, 2016 and January 8, 2016.
Quantity540
DistributionDistributed only in Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K902886CUSTOMED BASIC PACK III
Customed, Inc., Sep 19, 1990