Recall Z-1437-2016 ยท KKX Drape, Surgical, Antimicrobial
Customed, Inc
Catalog Number: 9001771 MAYO STAND COVER STERILE Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc.
| Date | Apr 14, 2016 (initiated Jan 7, 2016) |
| Status | Terminated |
| Reason | Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile. |
| Root cause | Packaging |
| Action | Recall Notification Letter sent to direct accounts via certified mail with return receipt on January 7, 2016 and January 8, 2016. |
| Quantity | 540 |
| Distribution | Distributed only in Puerto Rico. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K902886 | CUSTOMED BASIC PACK III Customed, Inc., Sep 19, 1990 |