Recall Z-0861-2016 ยท FPA Set, Administration, Intravascular

Centurion Medical Products Corporation

Convenience Kits from Centurion that contain an In-Vision- Plus Needleless IV Connector, Model: code RYM5001 InVision-Plus Needless IV Connectors are used for single patient use in IV and blood administration sets without the need for needles, thus eliminating the potential for needle-stick injuries during use.

DateFeb 23, 2016 (initiated Jan 19, 2016)
StatusTerminated
ReasonConvenience Kits contain an In-Vision-Plus Needleless IV Connector, code RYM5001 that is being recalled by the manufacturer, RyMed Technologies, LLC. A component molding abnormality may cause female luer threads to crack, leading to potential for the mating tubing set to become detached and for the administration to be interrupted. Other components within these kits are not affected.
Root causeNonconforming Material/Component
ActionThe firm, RyMed Technologies, LLC, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated 1/4/2016 to the affected customers via certified mail on 1/19/2016 with instructions for identifying and returning the affected product. The recall letter included a description of the reason for the recall, affected product, actions to be taken, and instructions for responding to the formal recall notification. The customers were instructed to discontinue distribution and/or use and return product to: RyMed Technologies, LLC, 6000 W. William Cannon Drive, Building B, Suite 300, Austin, TX 78749; that use
Quantity1130
DistributionUS Distribution in states of: TX, FL, VA, TN, PA, IL, GA, and CA.
FDA recordOpen on fda.gov