ClearanceBrief
Recall Z-1192-2016 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

QiG Group LLC

Algostim LLC, Algovita Implanted Pulse Generator stimulator, Model 2408 ( 3 x 8 channel)

DateMar 18, 2016 (initiated Jan 26, 2016)
StatusTerminated
ReasonThe implantable pulse generator (IPG) model 2408 had suboptimal performance establishing a connection to the external devices unless they were brought into close proximity (as close as 5 cm). The following functions may not be available: Reprogramming the stimulator, Adjusting stimulation parameters (on/off, intensity modifications, stimulation program selection), and Recharging the stimulator.
Root causeNonconforming Material/Component
ActionConsignee was sent on 1/26/2016 a Algostim, LLC "Urgent Field safety Notice" dated January 21, 2016. The letter described the problem and the product involved in the recall. The letter provided Background information as well as described the Clinical Implications, Rate of Occurrence, Recommendations and Mitigations. Advised consignee to deliver the respective letter to each physician who have implanted affected product and collect signed Response forms.
Quantity40
DistributionDistributed in Germany and Luxembourg.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.